New Model of Portable Prototype of IHLM
Note : 3 most unfortunate events occured in Test 5B.
(1) The Diaphragm Pump stops working during the test ; this DP was changed and the experiment continued.
(2) There was a leakage from the HALS which was subsequently controlled.
(3) The battery of the I-STAT machine of Jeetoo's Hospital was at low level ; the machine itself needs recalibration.
Ideal Heart Lung Machine - Test 5B Video
Ideal Heart Lung Machine - CPB Running (Duration : 35 secs)
ABG Results
ABG Results Table - Click to open/close
The table shows ABG results of 4 samples from I-STAT ABG machine at 0 min , 4 mins , 6 mins , and at 8 mins of circulation respectively.
Time of Circulation | 0 min | 4 mins | 6 mins | 8 mins |
pH | 6.951 | 6.979 | 6.948 | 6.948 |
PCO2 | 71.1 | 30.8 | 32.5 | 31.8 |
PaO2 | 70 | 242 | 144 | 133 |
BE | -16 | -24 | -25 | -25 |
HCO3 | 15.7 | 7.2 | 7.1 | 7.0 |
TCO2 | 18 | 8 | 8 | 8 |
SaO2 | 80% | 99% | 97% | 96% |
The Ideal Heart Lung Machine is the conclusion of the most fascinating history of Oxygenating Pumping Systems which started 2 centuries ago with Le Gallois in 1812.
Very severe gastroenteritis, bronchitis, extreme or very severe burns, very severe bronchial asthma.
Some auto immune diseases such as psoriasis, ulcerative colitis, rheumatoid arthritis and SLE(1 case only). I have planned, for a start, to apply a New Rhumatoid Arthritis Treatment Protocol and a New Psoriasis Treatment Protocol as soon as possible to patients suffering from RA including also RA patients who have not responded to DMARDs (Mauritius and abroad) and to Biologics Treatment Protocols (abroad).
The reason that I obtained much better results than other medical practitioners is that I have taken into consideration 2 most important aspects of the pathophysiologies of these diseases.
If one could substitute for the heart a kind of injection of artificial blood either naturally or artificially made, one could succeed in maintaining alive indefinitely any part of the body whatsoever.